SECTION I. IDENTIFICATION AND DESCRIPTION
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Signal Generator, Sg-644/U (Model 8614A) And 8616A Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . .
TB 9-6625-2224-24
Table 1. Calibration Description
Test instrument
parameters
Performance specifications
Frequency
SG-644/U (Hewlett-Packard, Model 8614A)
Range: 800 to 2400 MHz
Accuracy: 5 MHz
Line stability: 0.003% for 10% line voltage change
Hewlett-Packard, Model 8616A
Range: 1800 to 4500 MHz
Accuracy: 10 MHz
Stability: 0.003% for 10% line voltage change
RF output power
SG-644/U (Hewlett-Packard, Model 8614A)
Range: +10 to -127 dBm into 50  load
Accuracy: 0.75dB + attenuator accuracy (-10 to -127 dBm)1
(-15 to -127 dBm SN prefixed 501 and above)2
(0 to -127 dBm SN prefixed 851 and above)1
Hewlett-Packard, Model 8616A
Range: +10 to -127 dBm (1800 to 3000 MHz) into 50  load
+3 to - 127 dBm (3000 to 4500 MHz) into 50 load
Accuracy: 1.0 dB + attenuator accuracy (0 to -127 dBm)1
Attenuator
SG-644/U (Hewlett-Packard, Model 8614A)
Accuracy: +0, -3 dB (0 to -10 dBm) (0 to -15 dBm SN prefixed 501 and above)
.2 dB 0.06 dB/10 dB (-10 to -127 dBm)
(-15 to -127 dBm SN prefixed 501 and above)
Hewlett-Packard, Model 8616A
Accuracy: 0.2 dB 0.06 dB/10 dB (below -10 to -127 dBm)
Leveled output
SG-644/U (Hewlett-Packard, Model 8614A)
Range: Attenuator setting below 0 dB
Accuracy: .5 dB (.75 dB SN prefixed 851 and above)
Hewlett-Packard, Model 8616A
Range: Attenuator setting below 0 dB
Accuracy: 1.0 dB
Internal square wave
Range: 950 to 1050 Hz
(All models)
Verified at -10 dBm only.
1
Verified at -15 dBm only.
2
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, actual accuracy of the equipment selected is shown in parenthesis.