TB 9-6625-2227-24
Calibrator
Test instrument
RANGE
Test instrument indications
output
(dc)
pushbutton settings
Min
Max
190.0
mV
200
mV
189.6
mV
190.4
mV
1.900 V
2
V
1.897 V
1.903
V
190.0
V
200
V
189.7
V
190.3
V
1100
V
1200
V
1098
V
1102
V
Figure 1. Adjustments locations - models 3435A and 3438A.
a. Performance Check
(1) Connect calibrator OUTPUT to TI A and COM terminals.
mA and 200 A RANGE pushbuttons.
(2) Press FUNCTION
(3) Set calibrator output to 190 A.
(4) TI will indicate between 189.2 A and 190.7 A.
(5) Repeat technique of (2) and (3) above, using settings and indications listed in
table 4. TI will indicate within specified limits.