TB 9-6625-2227-24
Test instrument
RANGE
pushbutton
Calibrator output
Test instrument indications
settings
Voltage
Min
Max
200 mV
30 Hz
190
mV
186.9
mV
193.1
mV
100 Hz
190
mV
189.1
mV
190.9
mV
10 kHz
190
mV
189.1
mV
190.9
mV
100 kHz
190
mV
186.1
mV
193.9
mV
2V
30 Hz
1.9
V
1.869 V
1.931 V
100 Hz
1.9
V
1.891 V
1.909 V
10 kHz
1.9
V
1.891 V
1.909 V
100 kHz
1.9
V
1.861 V
1.939 V
200 V
30 Hz
190
V
186.9
V
193.1
V
100 Hz
190
V
189.1
V
190.9
V
10 kHz
190
V
189.1
V
190.9
V
100 kHz
190
V
186.1
V
193.9
V
1200 V
40 Hz
1000
V
982
V
1018
V
100 Hz
1000
V
994
V
1006
V
10 kHz
1000
V
994
V
1006
V
a. Performance Check
(1) Press FUNCTION k
and 20
RANGE pushbuttons.
and COM jacks. If TI does not indicate 0.00 , perform b (1) below,
(2) Short V
then remove short.
(3) Connect calibrator output terminals to TI INPUT V
and COM terminals.
(4) Press 20 k RANGE pushbutton and set calibrator to nominal 19 k . Adjust the
calibrator output adjustment control knob for a calibrator control display reading equal to the
TI indication. The calibrator control display ERROR indication will be within
.3000%; if
not, perform b (2) below.
(5) Set TI range and calibrator to the nominal resistance outputs listed in table 6.
At each resistance input, adjust the calibrator output adjustment control knob for a
calibrator control display reading equal to the TI indication. The calibrator control display
ERROR indication will be within the specified limits of table 6.
Test instrument
ERROR
Calibrator
RANGE
display indications
nominal
pushbutton
(%)
output
settings
1.5000
20
10
.4000
200
100
.4000
2k
1
k
.4000
200 k
100
k
.4000
2000 k
1
M
1.0000
20 M
10
M