TB 9-6625-2230-35
Test instrument
Calibrator
Calibrator error
range pushbutton
nominal
display indication
settings
settings
Min
Max
2 kΩ
1.000 kΩ
-.09%
+.09%
20 kΩ
10.000 kΩ
-.09%
+.09%
200 kΩ1
kΩ
-.09%
+.09%
100.00
MΩ
1.0000 MΩ
-.17%
+.17%
MΩ
MΩ
-.5%
+.5%
10.00
MΩ2
MΩ
MΩ
MΩ
50.0
49.2
50.8
MΩ3
MΩ
MΩ
MΩ
200
193
207
Set Calibrator 2 wire comp OFF.
1
2
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set TI to measure frequency by selecting ac voltage mode and pressing HZ
pushbutton.
(2) Press 200 mV pushbutton.
(3) Adjust calibrator output to 20 mV and frequency to 100 Hz. TI will indicate
between 99.94 and 100.06 Hz.
(4) Set calibrator output to 200 mV and frequency to 19.000 kHz. TI will indicate
between 18.990 and 19.010 kHz.
(5) Set calibrator output to 200 mV and frequency to 190.00 kHz. TI will indicate
between 189.90 and 190.10 kHz.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.