TB 9-6625-2230-35
a. Performance Check
(1) Connect calibrator output to TI A and COMMON terminals.
(2) Set TI to measure current on the 200 A range.
(3) Set calibrator output to 190.00 A. TI will indicate between 189.60 and 190.40
A.
(4) Repeat technique of (2) and (3) above, using settings and indications listed in
table 4. TI will
indicate within
limits specified.
b. Adjustments. No adjustments can be made.