TB 9-6625-2232-24
(e) MATH/REGISTERS STORE.
(f) MATH/REGISTERS 9/N DIG DISP.
(3) Set calibrator to settings listed in table 5. If TI indications are not within limits
specified, perform b below.
Calibrator output settings
Test instrument indications
Voltage (V)
Min
Max
1
25 Hz
0.995324
1.004676
1
1 kHz
0.997924
1.002076
1
45 kHz
0.995924
1.004076
1
95 kHz
0.990924
1.009076
1
250 kHz
0.942824
1.057176
10
25 Hz
9.95324
10.04676
10
1 kHz
9.97924
10.02076
10
45 kHz
9.95924
10.04076
10
95 kHz
9.90924
10.09076
10
250 kHz
9.42824
10.57176
100
25 Hz
99.5324
100.4676
100
1 kHz
99.7924
100.2076
100
45 kHz
99.5924
100.4076
100
95 kHz
99.0924
100.9076
700
1 kHz
697.8
702.2
700
45 kHz
696.1
703.9
700
95 kHz
692.24
707.76
b. Adjustments
NOTE
Adjustments are located behind a front panel section located at
the front input terminals.
(1) Set calibrator to STANDBY.
(2) Press TI RESET key then FUNCTION ~V key.
(3) Set calibrator for a 1 V, 1 kHz output.
key to 100 V range. Adjust K until TI indicates 01.000
(4) Press TI RANGE
or
1 count (R).
(5) Press TI RANGE
key to 1 V range. Adjust L until TI indicates 1.00000 3
counts (R).
(6) Press TI RANGE
key to 1 kV range. Adjust either M or N until TI indicates
001.00 2 counts (R).
(7) Repeat (4) through (6) above until no further adjustments are necessary.
(8) Set calibrator to STANDBY and disconnect from TI.
/4WR
HI and LO.
(9) Short TI VOLTS/2WR
8