Table 5. Ac Voltage

Calibration Procedure For Digital Voltmeter Model 3456A Page Navigation
  3    4    5    6    7  8
TB 9-6625-2232-24
+ = key and RANGE
key to 1 V range. Adjust N for
(10) Press FUNCTION
or
a minimum TI indication.
(11) Press FUNCTION ~V key and perform (a) through (c) below:
(a)  Record TI indication.
(b)  Adjust M for a minimum TI indication.
(c)  Continue adjusting M in the same direction until TI indication is within 2
counts of indication recorded in (a) above.
(12) Repeat (10) and (11) above until both indications are within 2 counts of each other.
11. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/(10 Blank)