SECTION II. EQUIPMENT REQUIREMENTS
Vertical Deflection

Calibration Procedure For Oscilloscope, An/Usm-425(V)1 (Type 465M) And Types 465, 465 With Options, 465B And 465B With Options Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . . . . .
TB 9-6625-2239-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
MULTIMETER
Range: 0 to +55 V
Fluke, Model 8840A/AF05
Accuracy: 0.25%
(AN/GSM-64D)
OSCILLOSCOPE
Voltage output:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range:
30 mV to 30 V
(5820A-5C-GHZ)
Frequency: 1 kHz
Accuracy:   0.25%
Time markers:
Range:
1 ns to 0.5 s
Accuracy:
.5%
Risetime:
0.875 ns (1 ns or less)
Aberrations:
0.875% (l %)
Level sine wave:
Reference frequency: 50 kHz
Amplitude: 30 mV
Capability: >100 MHz
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraphs 8 through 12 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 12.  Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel  fail  to  observe  safety  precautions.  REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.