Table 3. Trace Alignment
Figure 2. Tektronix, Type 465 and 465 with options - vertical preamp and attenuator circuit boards.

Calibration Procedure For Oscilloscope, An/Usm-425(V)1 (Type 465M) And Types 465, 465 With Options, 465B And 465B With Options Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. .
TB 9-6625-2239-24
Figure 1. AN/USM-425(V)1, (Tektronix, Type 465M) - vertical board.
Table 4. Vertical Deflection Accuracy
Test instrument
Oscilloscope
VOLTS/DIV
calibrator
switch settings
output voltage
10  m
60 mV
20  m
120 mV
50  m
0.3 V
0.1
0.6 V
0.2
1.2 V
0.5
3.0 V
1.0
6.0 V
2.0
12
V
5.0
30.0 V
7