TB 9-6625-2241-24
a. Performance Check
(1) Position controls as listed in (a) through (f) below:
A TIME/DIV switch to 1 ms.
(a)
B TIME/DIV switch to 5 μs.
(b)
HORIZ DISPLAY pushbutton A INTEN on (in).
(c)
DELAY TIME POSITION dial to 1.00.
(d)
VERT MODE pushbutton CH 1 to on (in).
(e)
Both VOLTS/DIV switches to .5 V.
(f)
(2) Connect oscilloscope calibrator to CH 1 input through 50 Ω termination.
(4) Adjust horizontal POSITION control to align 1st time marker with 1st vertical
graticule line. Adjust INTENSITY control to view intensified 2d time marker.
(5) Press HORIZ DISPLAY pushbutton B DLY'D to on (in). If start of signal is
not aligned with the start of the sweep trace, perform b (1) below.
(6) Press HORIZ DISPLAY pushbutton A INTEN to on (in).
(7) Set DELAY TIME POSITION dial to 9.00.
(8) Press HORIZ DISPLAY pushbutton B DLY'D to on (in). If start of signal is
not aligned with start of sweep trace, perform b (2) below.
(9) Press HORIZ DISPLAY pushbutton A to on (in) and set DELAY TIME
POSITION dial to full ccw.
(10) Adjust horizontal POSITION control to align 1st time marker with 1st vertical
graticule line. Adjust oscilloscope calibrator in EDIT mode for 1 marker per division. If
oscilloscope calibrator readout does not indicate within 2%, perform b (3) below.
(11) Press X10 MAG pushbutton to on (in).
(13) Repeat technique of (10) above.
If oscilloscope calibrator readout does not
indicate within 3%, perform b (4) below.
(14) Press and release X10 MAG pushbutton to off (out).
(16) Set B TRIGGER SOURCE switch to NORM.
(17) Set B TIME/DIV control to 1 ms and press HORIZ DISPLAY pushbutton B
DLY'D to on (in).
(18) Repeat technique of (10) above.
If oscilloscope calibrator readout does not
indicate within 2%, perform b (5) below.
(20) Position controls as listed in (a) through (e) below:
(a) DELAY TIME POSITION dial to 1.00.
(b) B TRIGGER SOURCE switch to STARTS AFTER DELAY.
8