Figure 1. Tektronix, Types 464, 466, and 466 opts 4/7 - overall view.
Table 5. B Sweep Timing

Calibration Procedure For Oscilloscope Types 464, 466, And 466 Options 4/7 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
TB 9-6625-2241-24
a. Performance Check
(1) Position controls as listed in (a) through (f) below:
A TIME/DIV switch to 1 ms.
(a)
B TIME/DIV switch to 5 μs.
(b)
HORIZ DISPLAY pushbutton A INTEN on (in).
(c)
DELAY TIME POSITION dial to 1.00.
(d)
VERT MODE pushbutton CH 1 to on (in).
(e)
Both VOLTS/DIV switches to .5 V.
(f)
(2) Connect oscilloscope calibrator to CH 1 input through 50 Ω termination.
(3) Set oscilloscope calibrator for MARKERS at 1 ms.
(4) Adjust horizontal POSITION control to align 1st time marker with 1st vertical
graticule line. Adjust INTENSITY control to view intensified 2d time marker.
(5) Press HORIZ DISPLAY pushbutton B DLY'D to on (in). If start of signal is
not aligned with the start of the sweep trace, perform b (1) below.
(6) Press HORIZ DISPLAY pushbutton A INTEN to on (in).
(7) Set DELAY TIME POSITION dial to 9.00.
(8) Press HORIZ DISPLAY pushbutton B DLY'D to on (in). If start of signal is
not aligned with start of sweep trace, perform b (2) below.
(9) Press HORIZ DISPLAY pushbutton A to on (in) and set DELAY TIME
POSITION dial to full ccw.
(10) Adjust horizontal POSITION control to align 1st time marker with 1st vertical
graticule line. Adjust oscilloscope calibrator in EDIT mode for 1 marker per division. If
oscilloscope calibrator readout does not indicate within 2%, perform b (3) below.
(11) Press X10 MAG pushbutton to on (in).
(12) Set oscilloscope calibrator for MARKERS at .1 ms.
(13) Repeat technique of (10) above.
If oscilloscope calibrator readout does not
indicate within 3%, perform b (4) below.
(14) Press and release X10 MAG pushbutton to off (out).
(15) Set oscilloscope calibrator for MARKERS at 1 ms.
(16) Set B TRIGGER SOURCE switch to NORM.
(17) Set B TIME/DIV control to 1 ms and press HORIZ DISPLAY pushbutton B
DLY'D to on (in).
(18) Repeat technique of (10) above.
If oscilloscope calibrator readout does not
indicate within 2%, perform b (5) below.
(19) Set oscilloscope calibrator for MARKERS at 5 μs.
(20) Position controls as listed in (a) through (e) below:
(a) DELAY TIME POSITION dial to 1.00.
(b) B TRIGGER SOURCE switch to STARTS AFTER DELAY.