TB 9-6625-2241-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
Range: 0.5 s/div to 0.05 μs/div
Delay time and differential
Accuracy: .01 major dial divisions for one major division or less
time measurement
1% between dial settings of more than one major dial division,
increasing to 0.75% over entire range.
Risetime
Range: 3.5 ns or less with aberrations not to exceed +3%, -3%, or a total
3% p-p (+4%, -4%, or a total 4% p-p for Tektronix, Type 466
SN B200000 and above)
Bandwidth
Dc to 100 MHz
Calibrator output
Range: 300 mV
Accuracy: 1% (0C to +40C)
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and the TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
5 - 80 pF standardizer.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
OSCILLOSCOPE
Voltage output:
Fluke, Model 5820A-5C-GHZ
CALIBRATOR
Range: 5 mV to 30 V
(5820A-5C-GHZ)
Frequency: 1 to 10 kHz
Accuracy: 0.75%
Time markers:
Range: 5 ns to 0.5 s
Accuracy: 0.5%
Risetime: 875 ns or less
Aberrations: .75%
Level sine wave:
Reference frequency: 50 kHz
Amplitude: 30 mV
Capacity: >100 MHz
MULTIMETER
Voltage input:
Fluke, Model 8840A/AF05
Range: 200 mV to 70 V dc
(AN/GSM-64D)
Accuracy: .25%
3