TB 9-6625-2243-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Multimeter, Fluke, Models 73, 73 Series III, 75, 75 Series II, 75 Series III, 77,
77 Series II, 77 Series III, 77/AN, and 77/BN. The manufacturers' manuals were used as
the prime data sources in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text and tables.
b. Time and Technique. The time required for each calibration is approximately 2 hours,
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which pertain
to this calibration are listed in table 1.
Models
Tables
73
73 Series III
75
75 Series II
75 Series III
77
77 Series II
77 Series III
77/AN
77/BN
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2