TB 9-6625-2246-40
f
d. Connect function/arbitrary generator Output to TI Input HI and LO using 50
eedthrough termination.
e. Set function/arbitrary generator for a sine wave, 100 kHz, 3 V rms output.
NOTE
In f through r below when SCAL command is executed
adjustment is automatically performed. When adjustment is
complete, TI returns to displaying dc voltage readings.
f. Execute SCAL 1E5 command by pressing keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift.
(2) MENU S/N Rdgs Trig.
(3) NUMERIC/USER 1E5,3458 then Enter.
g. Disconnect function/arbitrary generator and 50
feedthrough termination from TI.
h. Connect calibrator WIDEBAND to TI Input HI and LO using cable and
termination supplied with calibrator.
i. Set calibrator for a 100 kHz, 3 V wideband output.
j.
Execute SCAL 10 command by pressing keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift.
(2) MENU S/N Rdgs Trig.
(3) NUMERIC/USER 10,3458 then Enter.
k. Set calibrator for a 2 MHz, 3 V wideband output and repeat j above.
l.
Set calibrator for an 8 MHz, 3 V wideband output and repeat j above.
m. Set calibrator for a 100 kHz, 1 V wideband output.
n. Execute SCAL 1 command by pressing keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift.
(2) MENU S/N Rdgs Trig.
(3) NUMERIC/USER 1,3458 then Enter.
o. Set calibrator for an 8 MHz, 1 V wideband output and repeat n above.
p. Set calibrator for a 100 kHz, 100 mV wideband output.
q. Execute SCAL .1 command by pressing keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift.
(2) MENU S/N Rdgs Trig.
(3) NUMERIC/USER .1,3458 then Enter.
r. Set calibrator for an 8 MHz, 100 mV wideband output and repeat q above.
s. Press calibrator RESET key and disconnect from TI.
21