TB 9-6625-2246-40
b. Adjustments. If an out-of-tolerance condition exists, perform paragraph 18 below
and repeat 11 above.
a. Performance Check
(1) Press keys as listed in (a) through (f) below:
(a) FUNCTION/RANGE blue shift key then Reset/ key.
(b) FUNCTION/RANGE FREQ.
(c) FUNCTION/RANGE blue shift then MENU E/NPLC.
(d) FUNCTION/RANGE Menu Scroll until FSOURCE is displayed.
(e) NUMERIC/USER, then FUNCTION/RANGE Menu Scroll
until
ACDCV is displayed.
(f) NUMERIC/USER Enter.
(2) Connect function/arbitrary generator Output to TI Input HI and LO using
50
feedthrough termination.
(3) Set function/arbitrary generator for a sine wave, 20 Hz, 1 V rms output. TI will
indicate between 19.99000 and 20.01000 Hz.
(4) Set function/arbitrary generator frequency to 10 MHz. TI will indicate between
09.99900 and 10.00100 MHz.
b. Adjustments. If an out-of-tolerance condition exists, and if not previously
ADJUSTMENT PROCESS
13. Preliminary Instructions
a. Remove all external input signals from front and rear terminals.
b. Press FUNCTION/RANGE DCV key and Range
key to 100 mV range.
c. Release Terminals Front/Rear pushbutton to Front and Guard Open/To LO
pushbutton to Open.
d. Ensure that at least a 4 hour warm-up has elapsed since power was applied.
14. Front Terminal Offset
Sense and Input HI and LO as
a. Connect a 4 terminal short across front panel
b. After connecting short, allow 5 minutes for thermal stabilization.