TB 9-6625-2246-40
Test
instrument
Calibrator output settings
Test instrument indications
range settings
Voltage
Min
Max
10
mV
10
mV
15
Hz
9.928 mV
10.072
mV
10
mV
10
mV
30
Hz
9.960 mV
10.040
mV
10
mV
10
mV
90
Hz
9.969 mV
10.031
mV
10
mV
10
mV
15
kHz
9.973 mV
10.027
mV
10
mV
10
mV
45
kHz
9.960 mV
10.040
mV
10
mV
10
mV
90
kHz
9.895 mV
10.105
mV
10
mV
10
mV
200
kHz
9.530 mV
10.470
mV
100
mV
100
mV
90
kHz
99.32
mV
100.68
mV
100
mV
100
mV
200
kHz
97.50
mV
102.50
mV
100
mV
100
mV
450
kHz
96.40
mV
103.60
mV
100
mV
100
mV
950
kHz
93.00
mV
107.00
mV
1
V
1
V
90
Hz
0.9993 V
1.0007 V
1
V
1
V
45
kHz
0.9981 V
1.0019 V
10
V
10
V
15
Hz
9.9580 V
10.0420 V
10
V
10
V
30
Hz
9.9830 V
10.0170 V
100
V
100
V
15
Hz
99.5800 V
100.4200 V
100
V
100
V
30
Hz
99.8300 V
100.1700 V
100
V
100
V
90
Hz
99.9300 V
100.0700 V
100
V
100
V
15
kHz
99.9600 V
100.0400 V
100
V
100
V
45
kHz
99.8100 V
100.1900 V
100
V
100
V
90
kHz
99.4200 V
100.5800 V
100
V
20
V
200
kHz
19.1000 V
20.9000 V
100
V
20
V
450
kHz
18.8000 V
21.2000 V
100
V
20
V
950
kHz
17.0000 V
23.0000 V
1000
V
700
V
90
Hz
699.240 V
700.760 V
1000
V
700
V
15
kHz
699.380 V
700.620 V
1000
V
700
V
45
kHz
698.550 V
701.450 V
(6) Press calibrator RESET key and disconnect from TI.
(7) Connect calibrator WIDEBAND to TI Input HI and LO using cable and
termination supplied with calibrator.
(8) Press TI Guard Open/To LO pushbutton to To LO.
key to 100 mV range.
(9) Press FUNCTION/RANGE Range
(10) Set calibrator for a 100 mV, 2 MHz wideband output. TI will indicate between
85 and 115 mV.
(11) Press FUNCTION/RANGE Range
key to 1 V range.
(12) Set calibrator for a 1 V, 2 MHz wideband output. TI will indicate between 0.85
and 1.15 V.
(13) Press FUNCTION/RANGE Range
key to 10 V range.
(14) Set calibrator for a 3 V, 2 MHz wideband output. TI will indicate between 2.2
and 3.8 V.
17