TB 9-6625-2246-40
until TRIG AUTO is displayed.
(b) FUNCTION/RANGE Menu Scroll
(c) NUMERIC/USER Enter.
(8) Perform paragraphs 13 through 17 below on multimeter (not TI) to establish
manufacturer's 24 hour accuracy.
(10) Set multimeter to measure dc voltage on 1 V range.
or
key to 100 A range.
(11) Press TI FUNCTION/RANGE Range
(12) Set calibrator for a 100 A dc output, then using output adjustment controls;
adjust calibrator output for a 100.00000 (20 digits) TI indication.
(13) Divide multimeter indication by standard resistor test report value. Calculated
value will be between 99.99664 and 100.00336 A.
(14) Repeat technique of (9) through (13) above using standard resistors and settings
listed in table 6. Calculated values will be within limits specified.
15