Figure 6. Power supply - adjustment locations.

Calibration Procedure For Microwave Amplifier Model 8349B Page Navigation
  9    10    11    12    13  14
img
. . . . . .
TB 9-6625-2249-24
b. Adjustments
(1) Adjust A5R38 (fig. 6) for a multimeter indication between 4.999 and 5.001 V dc (R).
(2) Adjust A5R35 (fig. 6) for a multimeter indication between 7.999 and 8.001 V dc (R).
(3) Adjust A5R39 (fig. 6) for a multimeter indication between 14.999 and 15.001 V dc (R).
(4) Adjust A5R36 (fig. 6) for a multimeter indication between -14.999 and -15.001 V dc (R).
12. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
15/(16 Blank)