Table 10. Frequency Readout Accuracy 1.5 GHz
Table 12. Frequency Readout Accuracy 9 GHz

Calibration Procedure For Spectrum Analyzer An/Usm-489A Page Navigation
  18    19    20    21    22  23  24    25    26    27    28  
img
. . . . . . .
TB 9-6625-2250-24
MARKER PEAK SEARCH.
(b)
INSTRUMENT STATE MIXER INT.
(c)
[PRESEL AUTO PK] wait for PEAKING message to disappear.
(d)
SPAN then [SPAN WIDTH] to 1 MHz.
(e)
(9) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
within limits specified in first row of table 11.
(10) Repeat technique of (8)(e) and (9) above for remaining SPAN [SPAN WIDTH]
settings listed in table 11. TI MKR frequency indications will be within limits specified in
table 11.
Table 11. Frequency Readout Accuracy 4 GHz
Synthesized signal
Test instrument
generator
FREQUENCY
SPAN
MKR frequency indications
frequency
[SPAN WIDTH]
(CENTER FREQ]
(GHz)
setting
(GHz)
setting (GHz)
settings
Min
Max
4
4
1 MHz
3.999948
4.000051
4
4
10 MHz
3.99948
4.00051
4
4
20 MHz
3.99895
4.00104
4
4
50 MHz
3.99745
4.00254
4
4
100 MHz
3.9948
4.0051
4
4
1 GHz
3.950
4.050
(11) Set synthesized signal generator frequency to 9 GHz.
(12) Press TI keys and enter values using DATA keys as listed in (a) through (e) below:
FREQUENCY then [CENTER FREQ] to 9 GHz.
(a)
MARKER PEAK SEARCH.
(b)
INSTRUMENT STATE MIXER INT.
(c)
[PRESEL AUTO PK] wait for PEAKING message to disappear.
(d)
SPAN then [SPAN WIDTH] to 1 MHz.
(e)
(13) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
within limits specified in first row of table 12.
(14) Repeat technique of (12) (e) and (13) above for remaining SPAN [SPAN
WIDTH] settings listed in table 12. TI MKR frequency indications will be within limits
specified in table 12.