TB 9-6625-2250-35
EXAMPLE:
(a) If the IF GAIN DEVIATION values in table 5 are: [REF LVL] -50 dBm = +0.2 dB
[REF LVL] -40 dBm = -0.3 dB
(b) IF GAIN CORRECTION formula listed in table 6 is: (-50) - (-40) =
(c) Substitute table 5 IF GAIN DEVIATION values for [REF LVL] settings into
formula: (-50) - (-40) = (+0.2) - (-0.3) = +0.5
(d) Then +0.5 would be recorded as IF GAIN CORRECTION (dB) factor in table 6
for [ATTEN] 20 dB setting. NOTE: Do not record EXAMPLE in table 6.
Table 6. Input Attenuator Accuracy (18 GHz)
Test instrument
IF GAIN
Step-to-step
Corrected ∆ MKR
∆ MKR
[ATTEN]
CORRECTION
accuracy
settings
indications
indications
(dB)
(dB)
(dB)
(dB)
(dB)
10
0 (ref)
0 (ref)
0 (ref)
0 (ref)
20
(-50) - (-40) =
30
(-60) - (-40) =
40
(-70) - (-40) =
50
(-80) - (-40) =
60
(-90) - (-40) =
70
(-100) - (-40) =
(25) Repeat (24) above for each remaining IF GAIN CORRECTION listed in table 6.
Figure 3. Input attenuator accuracy (18 GHz) -
equipment setup.
(27) Press TI keys and enter values using DATA keys as listed in (a) through (d)
below:
(a)
(b)
AMPLITUDE then [REF LVL] to -10 dBm.
(c)
AMPLITUDE then [ATTEN] to 10 dB.
(d)
MARKER OFF.