TB 9-6625-2250-35
(28) Press synthesizer/level generator AMPLITUDE key.
(29) Press synthesizer/level generator INCR ⇓ key.
(30) Press TI AMPLITUDE key then [REF LVL] to -10 dBm.
(31) Press TI CONTROL TRIG key then [SINGLE] key and wait for completion of
sweep.
(32) TI ∆ MKR indication will be between 9 and 11 dB.
(33) Repeat technique of (29) through (31) above for synthesizer/level generator and
TI AMPLITUDE [REF LVL] settings listed in table 9. TI ∆ MKR indications will be
Synthesizer/level
Test instrument
generator amplitude
AMPLITUDE (REF LVL)
∆MKR indications
settings
settings
(dB)
(dBm)
(dBm)
Min
Max
-10
-20
-21
-19
-20
-30
-31
-29
-30
-40
-41
-39
-40
-50
-51
-49
-50
-60
-61
-59
-60
-70
-71
-69
-70
-80
-81
-79
a. Performance Check
(1) Connect signal generator RF Output to TI INPUT 50Ω.
(3) Press TI keys and enter values using DATA keys listed in (a) through (v) below:
(a)
INSTRUMENT STATE PRESET.
(b)
(c)
SPAN then [SPAN WIDTH] to 1 MHz.
(d)
AMPLITUDE then [REF LEVEL] to -10 dBm.
(e)
[LOG dB/DIV] to 1 dB.
(f)
CONTROL BW then [RES BW] to 3 kHz.
(g)
MARKER PEAK SEARCH.
(h)
[SIG TRK ON].
(i)
SPAN then [SPAN WIDTH] to 10 kHz.