Vertical Gain Accuracy
Table 9. Channel 1 Vertical Bandwidth Accuracy

Calibration Procedure For Oscilloscope Os-291/G Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . .
TB 9-6625-2251-24
Table 7. Channel 1 Vertical Gain Accuracy
Oscilloscope calibrator
CH1 VOLTS/DIV
VOLTAGE
err display
settings
settings
(%)
2
mV
8
mV
2.0
5
mV
20
mV
2.0
10
mV
40
mV
2.0
20
mV
80
mV
2.0
50
mV
200
mV
2.0
100
mV
400
mV
2.0
200
mV
800
mV
2.0
500
mV
2
V
2.0
1
V
4
V
2.0
2
V
8
V
2.0
5
V
20
V
2.0
i. Set oscilloscope calibrator to standby and select oscilloscope calibrator CHAN 2.
Select VERTICAL MODE, set CH 1 to off and set CH 2 to on.
Repeat technique of b, and e through g above for CH 2, with settings in table 8 below.
j.
Table 8. Channel 2 Vertical Gain Accuracy
Oscilloscope calibrator
CH2 VOLTS/DIV
VOLTAGE
err display
settings
settings
(%)
2
mV
8
mV
2.0
5
mV
20
mV
2.0
10
mV
40
mV
2.0
20
mV
80
mV
2.0
50
mV
200
mV
2.0
100
mV
400
mV
2.0
200
mV
800
mV
2.0
500
mV
2
V
2.0
1
V
4
V
2.0
2
V
8
V
2.0
5
V
20
V
2.0
k. Select VERTICAL MODE, set CH 1 to on and set CH 2 to off.
Set oscilloscope calibrator output to standby.
l.
10. Bandwidth Accuracy
NOTE
If TI does not perform as specified, perform adjustments as
indicated in SECTION IV.
7