TB 9-6625-2253-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Digital Multimeter, Fluke, Models 8024A and 8024B. The manufacturers'
manuals were used as the prime data sources in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Performance specifications
+(percent of reading plus digits)
Test instrument parameters
Dc voltage
Range: 0 to 1000 V in 5 ranges
Accuracy: r(0.1% +1)
Dc current
Range: 0 to 2000 mA in 4 ranges
Accuracy: r(0.75% +1)
Ac voltage
Range: 0 to 750 V in 5 ranges
Accuracy: 200 mV, 2, 20, 200 V ranges, 45 Hz to 1 kHz +(0.75% +2)
1 to 2 kHz (1.5% +3)
200 mV, 2, 20 V ranges 2 to 5 kHz r (5% +5)
750 V range 45 Hz to 1 kHz +(1% +2)
Ac current1
Range: 0 to 2000 mA in 4 ranges
Accuracy: 2 mA range, 45 to 450 Hz r(3% +2)
20 to 2000 mA ranges 45 Hz to 1 kHz +(1.5% +2)
Range: 0 to 20 M: in 6 ranges
Accuracy: 200: range r(0.2% +3) 2, 20 200 k: r (0.1% +1)
2000 k: r(0.15% +1)
20 M: r(2% +1)
Range: 200 nS
Accuracy: r(2.0% +10)
1Ac
current verified by dc current check.
Because current measurements of ac and dc are made using
same shunt resistors, a check of ac current is not made.