TB 9-6625-2253-24
(3) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. Calibrator Error display indication will be within
limits specified in first row of table 6.
(4) Repeat technique of (2) and (3) above, using the settings and indications listed in
table 6. Calibrator Error display indication will be within limits specified in table 6.
Table 6. Resistance Accuracy
Test instrument
Calibrator
Error indication
Range
Output
(%)
200
190.0  Ÿ
0.37
Ÿ
2  kŸ
1.9 kŸ
0.16
20  kŸ
19
kŸ
0.16
1
200  kŸ
190
kŸ
0.16
2000  kŸ
1.9 MŸ
0.21
20 MŸ
10  MŸ
2.1
2-wire comp: OFF
1
(5) Set TI to 200 nS range and set calibrator output to 10M: nominal. TI will
indicate between 97.0 and 103.0 nS.
b. Adjustments. No adjustments can be made.
12. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
CHANGE 1 7/(8 Blank)