TB 9-6625-2254-24
(8) Adjust the calibrator for 190.00 A dc. If TI does not indicate within limits
specified in first row of table 8, perform b (2) below.
(9) Repeat technique of (7) and (8) above for indications and settings listed in
table 8. If TI does not indicate within limits specified in table 8, perform b (2) below.
Calibrator
Test instrument
Indication limits
Output
DCA range
Min
Max
190.00
A
200 A
189.79 A
190.21 A
1.9000
mA
2 mA
1.8979 mA
1.9021 mA
19.000
mA
20 mA
18.979 mA
19.021 mA
190.00
mA
200 mA
189.79 mA
190.21 mA
1900.0
mA
2000 mA
1897.9
mA
1902.1
mA
b. Adjustments
(1) Simultaneously press the
(up) and
(down) pushbuttons.
After the TI
indicates "doNE", the display will indicate zero l count.
(2) With the appropriate current (determined by the select range) applied,
simultaneously press the (up) or (down) pushbuttons. After the TI indicates "doNE",
the display will indicate the applied current 1 count (R).
a. Performance Check
NOTE
Do not exceed 1000 V peak or 500 V rms between the input
V/ /TEMP and LOW terminals or instrument damage may
occur.
(1) Press ACV pushbutton.
(2) Connect TI V/ /TEMP and LOW terminals to the calibrator HI and LO
output terminals.
(3) Press
(up) or
(down) pushbutton to select the 200 mV range.
(4) Adjust the calibrator for 10.000 mV at 20 Hz. If TI does not indicate within
limits specified in first row of table 9, perform b below.
(5) Repeat technique of (3) and (4) above for voltages and settings in table 9. If TI
does not indicate within limits specified in table 9, perform b below.
Indication limits (18 C to 28 C)
DMM ACV
Applied ac
range
voltage
1
20 Hz
110 Hz
300 Hz
200
mV
10.000 mV
9.820 to 10.180
9.870 to 10.100
9.870 to 10.130
200
mV
190.000 mV
188.38 to 191.62
189.33 to 190.67 189.33 to 190.67
2
V
0.1000 V
0.0982 to 0.1018
0.0987 to 0.1013 0.0987 to 0.1013
See footnote at end of table.