TB 9-6625-2254-24
Table 9. Ac Voltage - Continued
Indication limits (18 C to 28 C)
DMM ACV
Applied ac
range
voltage
1
20 Hz
110 Hz
300 Hz
20 V
1.000 V
0.982 to 1.018
0.987 to 1.013
0.987 to 1.013
20 V
19.0000 V
18.838 to 19.162
18.933 to 19.067 18.933 to 19.067
200 V
10.000 V
9.82 to 10.18
9.87 to 1.013
9.87 to 10.13
200 V
190.000 V
188.38 to 191.62
189.33 to 190.67 189.33 to 190.67
500 V
100.00
V
98.9 to 101.1
99.4 to 100.5
99.4 to 100.5
500 V
500.00
V
495.7 to 504.3
498.2 to 501.8
498.2 to 501.8
DMM ACV
Applied ac
1 kHz
15 kHz
50 kHz
range
voltage
200 mV
10.000 mV 9.870 to 10.130
9.840 to 10.160
9.800 to 10.200
200 mV
190.000 mV
189.33 to 190.67
188.76 to 191.24 188.00 to 192.00
2V
.1000 V
0.0987 to 0.1013
0.0984 to 0.1014 0.0980 to 0.1020
2V
1.90000 V
1.8933 to 1.9067
1.8876 to 1.9124 1.8800 to 1.9200
20 V
1.000 V
0.987 to 1.013
0.984 to 1.016
0.980 to 1.020
20 V
19.0000 V
18.933 to 19.067
18.876 to 19.124 18.800 to 19.200
200 V
10.000 V
9.87 to 10.13
9.84 to 10.16
9.80 to 10.20
200 V
190.000 V
189.33 to 190.67
188.76 to 191.24 188.00 to 192.00
500 V
100.00
V
99.4 to 100.5
99.2 to 100.8
98.7 to 101.3
500 V
500.00
V
498.2 to 501.8
496.7 to 503.3
494.7 to 505.3
Check at 50 Hz when substituting an ac calibrator not specified below 50 Hz.
1
b. Adjustments
(1) With the appropriate voltage (5 percent of selected range) and 300 Hz frequency
applied, simultaneously press the (up) and (down) pushbuttons.
(2) After the TI indicates "190?" adjust calibrator to apply 95% of selected range
and 300 Hz frequency then simultaneously press the
(up) and
(down) pushbuttons.
After the TI indicates "doNE" the display will indicate the applied voltage l count (R).
a. Performance Check
NOTE
Do not exceed 2A at the mA and LOW input terminals.
(1) Press ACA pushbutton.
(2) C o n n e c t T I m A a n d L O W t e r m i n a l s t o t h e c a l i b r a t o r H I a n d L O o u t p u t
terminals.
(down) pushbutton to select the 200 A range.
(3) Press
(up) or
(4) Adjust the calibrator for 10.000 A at 20 Hz. If TI does not indicate within
limits specified in first row of table 10, perform b below.
(5) Repeat technique of (3) and (4) above for currents, frequencies, and settings in