TB 9-6625-2258-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Residual FM
Frequency range: 10 MHz to 4 GHz
Residual FM: Less than 5 kHz peak
Frequency range: 4 to 6.5 GHz
Residual FM: Less than 7.0 kHz peak
Frequency range: 6.5 to 20 GHz1
Residual FM: Less than 15 kHz peak
Frequency range: 20 to 40 GHz1
Residual FM: Less than 30.0 kHz peak
Output maximum power
Frequency range: 10 MHz to 2 GHz
Output power: At least +13 dBm
Frequency range: 2 to 18 GHz
Output power: At least +10 dBm
Frequency range: 18 to 20 GHz
Output power: At least +7 dBm
Frequency range: 20 to 40 GHz
Output power: At least 0 dBm
Output attenuator
Frequency range: 10 MHz to 2 GHz
Attenuator range: 0 to -60 dBm
Total maximum error: 2.6 dB
Attenuator range: -60 to -80 dBm
Total maximum error: 3.1 dB
Frequency range: 2 to 20 GHz1
Attenuator range: 0 to -60 dBm
Total maximum error: 2.6 dB
1Not
checked above 18 GHz due to standards limitations.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287, AN/GSM-705 and Secondary Reference Calibration Standards
Set NSN 4931-00-621-7878. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the actual accuracy of the
equipment selected is shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.