(12) Configure measuring receiver to measure frequency.
(13) Remove TI top cover.
(14) Adjust A6R12 (fig. 1) for a measuring receiver frequency indication between
2100 and 2102 MHz (R).
Figure 1. A6 board - component locations.
(15) Replace TI top cover and wait 5 minutes before proceeding to (16) below.
(16) Press FREQUENCY RANGE CW F2 key.
(17) Configure measuring receiver to measure frequency.
(18) Remove TI top cover.
(19) Adjust A6R6 (fig. 1) for a measuring receiver frequency indication between 7898
and 7900 MHz (R).
(20) Replace TI top cover and wait 5 minutes before proceeding to (21) below.
(21) Repeat (8) through (20) above until both FREQUENCY RANGE CW F1 and
CW F2 frequencies are within the specified limits.