TB 9-6625-2258-24
(78) Replace TI top cover.
(79) Press FREQUENCY RANGE CW F2 key and wait 10 seconds for frequency
count to settle.
(80) Remove TI top cover.
(81) Adjust A9R6 (fig. 2) for a microwave frequency counter indication between
27.398 and 27.400 GHz (R).
(82) Repeat (72) through (81) above until both F1 and F2 frequencies are within the
specified limits.
(83) Press RESET key.
(84) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to 27.600
GHz by pressing 27.6 with DATA ENTRY keys and then by pressing GHz/dBm/SEC
DATA ENTRY key.
(85) Press FREQUENCY RANGE CW F2 key and set F2 M6 display to 40.000
GHz by pressing 40 with DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA
ENTRY key.
(86) Press FREQUENCY RANGE CW F1 key and wait 10 seconds for frequency to settle.
(87) Press FREQUENCY RANGE CW F2 key and wait 10 seconds for frequency to settle.
(88) Repeat (86) and (87) above two more times.
(89) Press FREQUENCY RANGE CW F1 key and wait 10 seconds for frequency
count to settle.
(90) Remove TI top cover.
(91) Adjust A8R12 (fig. 3) for a microwave frequency counter indication between
27.600 and 27.602 GHz (R).
(92) Replace TI top cover.
(93) Press FREQUENCY RANGE CW F2 key and wait 10 seconds for frequency
count to settle.
(94) Remove TI top cover.
(95) Adjust A8R6 (fig. 3) for a microwave frequency counter indication between
39.998 and 40.002 GHz (R).
(96) Replace TI top cover.
(97) Repeat (86) through (96) above until both CW F1 and CW F2 frequencies are
within the specified limits.
(98) Press POWER pushbutton to OFF position and remove TI top cover.
(99) Replace screw holding CCA boards.
13