TB 9-6625-2258-24
(4) Press F2 M6 key and set F2 M6 display to 40.000 GHz by pressing 40 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(5) Press LEVEL key and set LEVEL display to -10 dBm by pressing -10 with
DATA ENTRY keys and then by pressing GHz/dBm/ SEC DATA ENTRY key.
(6) Press TRIGGER MANUAL SWEEP key.
(7) Connect multimeter (dc mode) to V/GHz OUTPUT (rear panel).
to INPUT 1 of microwave frequency counter.
(8) Connect RF OUTPUT 50
(9) Adjust TRIGGER MANUAL SWEEP control for a microwave frequency
counter indication on 40.000 GHz.
(10) Remove TI top cover.
(11) Adjust A1A5R47 (located on A1A5 board) for a multimeter indication between
19.98 and 20.02 V dc.
(12) Replace TI top cover.
connection from INPUT 1 to INPUT 2 of microwave
(13) Move RF OUTPUT 50
frequency counter.
(14) Adjust TRIGGER MANUAL SWEEP control for microwave frequency counter
(15) Record multimeter indication in table 6.
(16) Move RF OUTPUT 50W connection from INPUT 2 to INPUT 1 of microwave
frequency counter.
(17) Repeat (14) and (15) above for remaining microwave frequency counter
Test instrument
Test instrument
Microwave
Microwave
V/GHz OUTPUT
V/GHz OUTPUT
frequency
frequency
multimeter indications
multimeter indications
counter
counter
(V dc)
(V dc)
indications
indications
(GHz)
(GHz)
0.0101
21.000
1.000
22.000
2.000
23.000
3.000
24.000
4.000
25.000
5.000
26.000
6.000
26.500
7.000
27.000
8.000
28.000
9.000
29.000
10.000
30.000
11.000
31.000
12.000
32.000
13.000
33.000
14.000
34.000
18