Table 4. Residual AM Test Frequencies
Table 6. V/GHz Output Check

Calibration Procedure For Programmable Sweep Generator, Model 6669M Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . . . . . .
TB 9-6625-2258-24
12. Residual FM Test
a. Performance Check
(1) Press TI RESET key.
(2) Configure measuring receiver to measure FM modulation with <20 Hz high-pass
filter, 15 kHz low-pass filter and RMS detector selected.
(3) Press LEVEL key and set LEVEL display to +13.0 dBm by pressing 13 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(4) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to 0.010 GHz
by pressing 10 with DATA ENTRY keys and then by pressing MHz/dB/mS DATA
ENTRY key.
(5) Measuring receiver FM modulation display will indicate less than 5.0 kHz.
(6) Repeat technique of (3) through (5) above for 0.100, 1.0, and 1.7 GHz.
(7) Press LEVEL key and set LEVEL display to +10.0 dBm by pressing 10 with
DATA ENTRY keys and then by pressing GHZ/dBm/SEC DATA ENTRY key.
(8) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to F1 M3
display frequency indications listed in table 5.
(9) Measuring receiver FM modulation display will indicate less than value listed in
table 5.
(10) Repeat technique of (7) through (9) above for remaining F1 M3 display
frequency indications listed in table 5.
Table 5. Residual FM Test
Measuring
Test instrument
receiver
F1 M3 display
MODULATION
frequency
display
indications
(kHz)
(GHz)
3.000
<5.0
4.000
<5.0
5.000
<7.0
6.000
<7.0
12.000
<15.0
17.500
<15.0
b. Adjustments. No adjustments can be made.
13. V/GHz Output Test
a. Performance Check
(1) Press RESET key.
(2) Press FREQUENCY RANGE Fl-F2 key.
(3) Press Fl M3 key and set F1 M3 display to 0.010 GHz by pressing 10 with
DATA ENTRY keys and then by pressing MHz/ dB/mS DATA ENTRY key.