• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Spurious Signal Test
Residual FM Test

Calibration Procedure For Programmable Sweep Generator, Model 6669M Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . . .
TB 9-6625-2258-24
(10) Slowly rotate INCR-DECR control cw (from 2.000 to 18.000 GHz) and observe
non-harmonic related (spurious) signals. Spurious signals will be <-45 dBc (45 dB below
carrier level indication on spectrum analyzer).
b. Adjustments. No adjustments can be made.
11. Residual AM Test
a. Performance Check
(1) Press TI RESET key.
(2) Connect measuring receiver power sensor to TI RF OUTPUT 50 .
(3) Configure measuring receiver to measure AM modulation with <20 Hz high-pass
filter, 300 kHz low-pass filter and RMS detector selected.
(4) Press LEVEL key and set LEVEL display to +13.0 dBm by pressing 13 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(5) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to 0.010 GHz
by pressing 10 with DATA ENTRY keys and then by pressing MHz/dB/mS DATA
ENTRY key.
(6) Measuring receiver AM modulation display will indicate less than 0.32 percent.
(7) Repeat technique of (4) through (6) above for 0.100, 1.0, and 1.7 GHz.
(8) Press LEVEL key and set LEVEL display to +10.0 dBm by pressing 10 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(9) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to F1 M3
display frequency indication listed in table 4.
(10) Measuring receiver AM modulation display will indicate less than 0.32 percent.
(11) Repeat technique of (8) through (10) above for remaining F1 M3 display
frequency indications listed in table 4.
Table 4. Residual AM Test Frequencies
Test instrument F1 M3
display frequency
indications
(GHz)
3.000
5.000
10.000
15.000
17.500
b. Adjustments. No adjustments can be made.