TB 9-6625-2265-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the
calibration of Digital Multimeter, Ballantine, Model 3100A. The manufacturer's manual
was used as the prime data source in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations . None.
b. Time and Technique . The time
required for this
calibration is
approximately 1
hour using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for
calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications (3 1/2 digit display)
parameters
Dc voltage
Range: 0 to 1000 V (in 5 ranges)
Accuracy: (0.1% of reading + 1 digit)
Ac voltage
Range: 0 to 750 V (in 5 ranges)
Accuracy: (% of reading + digits)
Range
Frequency
40 Hzto 1 kHz
1 to 2 kHz
2 to 5 kHz
200
mV
.75 + 2
1.5 + 3
5.0 + 5
2
V
20
V
200
V
N/A
750
V
1.0 + 2
N/A
N/A
Range: 0 to 20 MΩ (in 6 ranges)
Accuracy: (% of reading + digits)
Range:
200Ω ...............................................................
0.2
+
3
2 through 200 kΩ .........................................
0.1
+
1
2 MΩ ...............................................................
.15
+
1
20 MΩ .............................................................
2.0
+
1