TB 9-6625-2267-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc voltage range: -190 mV to 1000 V
Fluke, Model 5720A
Voltage:
Accuracy: (%)
(5720A) (p/o MIS-35947);
190 mV ..................................................
0.013
w amplifier, Fluke
-1.9 to 190 V ............................................
0.0066
5725A/AR (5725A/AR)
1000 V .....................................................
0.0075
Ac voltage:
Range: 190 mV to 1000 V
Accuracy: (%)
Voltage
0.030
5.0
0.040/20
50
100
190 mV
0.151
0.071
N/A
0.151
0.257
1.9 V through
0.132
0.0541
N/A
0.132
0.263
190 V
400 V
N/A
0.075
N/A
N/A
N/A
1000 V
N/A
0.102
0.150
N/A
N/A
Dc current range: 190 A to 1.9 A
Current:
Accuracy: (%)
190 A to 190 mA ...................................
0.028
1.9 A ........................................................
0.053
Accuracy: (%)
100 .......................................................
0.0325
1.0 k ......................................................
0.0275
10, 100, and 1000 k ..............................
0.0150
10 M ......................................................
0.0525
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.