TB 9-6625-2267-24
a. Performance Check
and LO.
(1) Connect calibrator OUTPUT HI and LO to TI INPUT V-
(2) Press FUNCTION ACV pushbutton.
(3) Press TI RANGE pushbuttons and set calibrator output for settings listed in
table 5. If TI indications are not within limits specified, perform b below.
Test instrument
Calibrator
Test instrument
RANGE
output settings
indications
pushbuttons
Voltage
Min
Max
200 mV
190
mV
30 Hz
188.85
mV
191.15
mV
200 mV
190
mV
5 kHz
189.46
mV
190.54
mV
200 mV
190
mV
50 kHz
188.85
mV
191.15
mV
200 mV
190
mV
100 kHz
188.05
mV
191.95
mV
2
1.9 V
30 Hz
1.8900 V
1.9100
V
2
1.9 V
500 Hz
1.8959 V
1.9041
V
2
1.9 V
5 kHz
1.8959 V
1.9041
V
2
1.9 V
50 kHz
1.8900 V
1.9100
V
2
1.9 V
100 kHz
1.8800 V
1.9200
V
20
19
V
30 Hz
18.900
V
19.100
V
20
19
V
5 kHz
18.959
V
19.041
V
20
19
V
50 kHz
18.900
V
19.100
V
20
19
V
100 kHz
18.800
V
19.200
V
200
190
V
30 Hz
189.00
V
191.00
V
200
190
V
5 kHz
189.59
V
190.41
V
200
190
V
50 kHz
189.00
V
191.00
V
200
190
V
100 kHz
188.00
V
192.00
V
1200 V
400
V
5 kHz
398.8
V
401.2
V
1200 V
1000
V
40 Hz
994.0
V
1006.0
V
1200 V
1000
V
5 kHz
995.9
V
1004.1
V
1200 V
1000
V
20 kHz
994.0
V
1006.0
V
b. Adjustments
(1) Set calibrator to STANDBY.
(2) Press RANGE 1200V pushbutton.
(3) Simultaneously press FUNCTION AC MA and DCV pushbuttons. Adjust R72
(fig. 1) until TI indicates between -20 and -30.
(4) Press FUNCTION ACV and RANGE 2 pushbuttons.
(5) Set calibrator for a 1.9 V, 500 Hz output. Adjust R50 (fig. 1) until TI indicates
1.9000 5 digits (R).
(6) Press RANGE 200 pushbutton.
(7) Set calibrator for a 190 V, 50 kHz output. Adjust C2 (fig. 1) until TI indicates
190.00 5 digits (R).