TB 9-6625-2268-24
Test
Calibrator
instrument
Error
range pushbutton
indications
(%)
settings
Output settings
1
0.121
200
190
0.111
2
1.9 k
0.061
20
19
k
0.061
2002
190
k
0.266
2000
1.9 M
0.266
20 M
19
M
1
Set calibrator 2 wire Comp to ON.
2
Set calibrator 2 wire Comp to OFF.
and 2 range pushbuttons to select 2 mS range.
(4) Simultaneously press 200
(5) Set calibrator for a 1 k
output.
(6) Divide TI indication into 1 and record results after rounding to 5 digits. Use
calibrator output adjustment controls to set calibrator control display Reading equal to
recorded value. Calibrator control display Error indication will be within 0.15%.
(7) Simultaneously press 2000 and 20 M
range pushbuttons to select
200 nS range.
(8) Set calibrator for a 10 M output and repeat technique of (6) above. Calibrator
control display Error indication will be within 0.7%.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/(10 Blank)