Figure 1. Adjustment locations.
Dc Current

Calibration Procedure For Digital Multimeter, Models 8050A And 8050A-01 Page Navigation
  1    2    3    4    5  6  7    8  
img
. . . . .
TB 9-6625-2268-24
(2) Press AC/DC pushbutton to AC.
(3) Press TI range pushbutton and set calibrator output to settings listed in table 4.
If TI indications are not within limits specified, perform b below.
Table 4. Ac Voltage
Test instrument
Calibrator output
Test instrument
range
settings
indications
pushbutton
settings
Voltage
Frequency
Min
Max
200
mV
190
mV
30  Hz
188.00
192.00
200
mV
190
mV
5  kHz
188.95
191.05
200
mV
190
mV
15  kHz
188.00
192.00
200
mV
190
mV
50  kHz
180.20
199.80
2
1.9  V
30  Hz
1.8800
1.9200
2
1.9  V
5  kHz
1.8895
1.9105
2
1.9  V
15  kHz
1.8800
1.9200
2
1.9  V
50  kHz
1.8020
1.9980
20
19
V
30  Hz
18.800
19.200
20
19
V
5  kHz
18.895
19.105
20
19
V
15  kHz
18.800
19.200
20
19
V
50  kHz
18.020
19.980
200
190
V
30  Hz
188.00
192.00
200
190
V
5  kHz
188.95
191.05
200
190
V
15  kHz
188.00
192.00
200
190
V
50  kHz
180.20
199.80
750
V ac
750
V
40  Hz
741.5
758.5
750
V ac
750
V
1  kHz
745.3
754.7
b. Adjustments
(1) Press TI 2 range pushbutton and set calibrator for a 1.9 V, 45 Hz output.
Adjust R7 (fig.1) for a TI indication of 1.9000 (5 digits). (R)
(2) Set calibrator for a 100 mV, 45 Hz output. Adjust R29 (fig. 1) for a TI indication
of .1000 (l digit). (R)
(3) Repeat (1) and (2) above until no further adjustments are required.
(4) Press TI 20 range pushbutton and set calibrator for a 19 V, 10 kHz output.
Adjust C1 (fig. 1) for a TI indication of 19.000 (10 digits). (R)
(5) Press TI 200 range pushbutton and set calibrator for a 100 V, 10 kHz output.
Adjust C2 (fig. 1) for a TI indication of 100.00 (5 digits). (R)
(6) Repeat (4) and (5) above until no further adjustments are required.
10. dB Display
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI V/k /S and COMMON.
(2) Simultaneously press V and mA function pushbuttons to select dB function.
(3) Press TI range pushbutton and set calibrator output to settings listed in table 5.
TI indications will be within limits specified.