TB 9-6625-2268-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications
parameters
(4 1/2 digit display)
dB display
Range: 0.77 mV to 750 V (-60 to +60 dBm with 600 reference)
Accuracy: (dBm)
Input
dBm
Range
0.020 0 .045 1.0 10
20
50
0.77
to2.0mV
-60 to -52
200 mV
0.5
N/A
2.0mV to2.0 V
-52 to +8
200 mV
0.01
to2.0 V
-18 to +8
2V
0.25
0.15
0.25
0.75
1.0
to20 V
+2 to +28
20 V
10
to200V
+22 to +48
200 V
100
to750V
+42 to +60
750 V
NOT SPECIFIED
Ac current2
Range: 0 to 2000 mA (in 5 ranges)
Frequency: 20 Hz to 20 kHz
Accuracy: (% of reading + digits)
Range
Frequency
20 Hz 45Hz 2kHz 10kHz
20kHz
200 A through 200 mA
2 + 10
1 + 10
2 + 10
2000 mA
NOT SPECIFIED
Dc current
Range: 0 to 2000 mA (in 5 ranges)
Accuracy: (0.3% of reading + 2 digits)
Range: 0 to 20 M (in 6 ranges)
Accuracy: (% of reading + digits)
200 and 2 k ............... 0.1 + 2 + 0.02
20 and 200 k ............... 0.05 + 2
2000 k and 20 M ...... 0.25 + 3
Range:
Accuracy:
(0.1% of reading + 5 digits)
2 ms
(0.5% of reading + 20 digits)
200 ns
1
Volts/hertz product not to exceed 107.
2
Ac current verified during dc current check since same shunt resistors are utilized for both functions.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this