TB 9-6625-2268-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Digital Multimeter, Fluke, Models 8050A and 8050A-01. The manufacturer's
manual was used as the prime data source in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Model 8050A-01 has rechargeable batteries; otherwise, both
models are the same.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with . (R)
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Performance specifications
Test instrument
parameters
(4 1/2 digit display)
Range: 0 to 1000 V (in 5 ranges)
Accuracy: (.03% of reading + 2 digits)
Dc voltage
Range: 0 to 750 V (in 5 ranges)
Accuracy: (% of reading + digits)
Ac voltage
Range
Frequency
20 Hz
45 Hz
1 kHz
10 kHz
20 kHz
50 kHz
200 mV through 200 V
1.0 + 10
.5 + 10
1.0 + 10
5.0 + 30
750 V
NOT SPECIFIED
See footnotes at end of table.