Calibration Procedure For Digital Multimeter, Models 8021B, 8022A And 8022B Page Navigation
  1    2  3  4  
img
. . . . . . . .
TB 9-6625-2269-24
Figure 1. Adjustment locations.
9. AC Voltage
a. Performance Check
(1) Press DC/AC pushbutton to AC (in) and press the 200 mV range pushbutton.
(2) Set TI range and calibrator for voltages and frequencies listed in table 4. TI will
indicate within the specified limits; if not, perform b below.
b. Adjustments. Set TI range to 200 mV and calibrator for a 190 mV, 45 Hz output.
Adjust AC CAL R4 (fig. 1) for a TI indication of 190.0 (R).
Table 4. Ac Voltage
Test instrument
Calibration
Test instrument
range pushbutton
output
indications
settings
Voltage
Frequency (Hz)
Min
Max
200 mV
190  mV
45
187.8
192.2
200 mV
190  mV
450
187.8
192.2
2
1.9  V
45
1.878
1.922
2
1.9  V
450
1.878
1.922
20
19  V
45
18.78
19.22
20
19  V
450
18.78
19.22
200
190  V
45
187.8
192.2
200
190  V
450
187.8
192.2
750 AC
750  V
45
740
760
750 AC
750  V
450
740
760
10. DC Current
a. Performance Check
(1) Connect calibrator OUTPUT terminals to TI mA and COMMON terminals, set
TI range to 2 mA, and set DC/AC pushbutton to DC (out).
(2) Set TI range and calibrator output as specified in table 5. TI will indicate within
the specified limits.
b. Adjustments. No adjustments can be made.