Figure 1. Adjustment locations.

Calibration Procedure For Digital Multimeter, Models 8021B, 8022A And 8022B Page Navigation
  1    2    3  4
img
. . . . . .
TB 9-6625-2269-24
Table 5. Dc Current
Test instrument
Test instrument
range pushbutton
Calibrator output
indications
settings
(dc current)
Min
Max
2 mA
1.9  mA
1.885
1.915
20 mA
19
mA
18.85
19.15
200 mA
190
mA
188.5
191.5
2000 mA
1.9  A
1885
1915
11. Resistance
a. Performance Check
and COMMON terminals. Set
(1) Connect calibrator OUTPUT terminals to TI
TI range to 200  and function pushbutton to  .
(2) Set TI range pushbutton and calibrator output to the settings listed in table 6.
The TI will indicate within the limits specified.
Table 6. Resistance
Test instrument
Test instrument
range pushbutton
Calibrator output
indications
settings
(resistance)
Min
Max
190
189.1
190.9
200
1.895
19.05
2k
1.9 k
18.95
19.05
20 k
19 k
189.5
190.5
200 k
190 k
1895
1905
2000 k
1.9 M
10 M
9.80
10.20
20 M
12. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.