SECTION III. CALIBRATION PROCESS
Ac Voltage

Calibration Procedure For Digital Multimeter An/Usm-486 (U) (Model 8050A/Fm) Page Navigation
  1    2    3  4  5    6    7  
img
. . . . . .
TB 9-6625-2270-24
Table 3. Dc Voltage
Test instrument
Calibrator output
Test instrument indications
range pushbutton
(V dc)
Adjustments
Min
Max
200
mV
0.19
189.88
190.12
b(l)
200
mV
-0.19
-189.88
-190.12
-----
2
1.9
1.8988
1.9012
b(2)
20
19
18.988
19.012
-----
200
190
189.88
190.12
b(3)
1000
DC
1000
999.2
1000.8
b(4)
b. Adjustments
(1) Adjust R12 (fig. 1) for a TI indication of 190.00 (R).
(2) Adjust R11 (fig. 1) for a TI indication of 1.9000 (R).
(3) Adjust R5 (fig. 1) for a TI indication of 190.00 (R).
(4) Adjust R6 (fig. 1) for a TI indication of 1000.0 (R).
Figure 1. Adjustment locations.