TB 9-6625-2270-35
Test instrument indications
Test instrument
Calibrator output
range pushbutton
(V dc)
Min
Max
Adjustments
200 mV
.19
189.88
190.12
b(l)
200 mV
-.19
-189.88
-190.12
2
1.9
1.8988
1.9012
b(2)
20
19
18.988
19.012
200
190
189.88
190.12
b(3)
1000 DC
1000
999.2
1000.8
b(4)
b. Adjustments
(1) Adjust R12 (fig. 1) for a TI indication of 190.00 (R).
(2) Adjust R11 (fig. 1) for a TI indication of 1.9000 (R).
(3) Adjust R5 (fig. 1) for a TI indication of 190.00 (R).
(4) Adjust R6 (fig. 1) for a TI indication of 1000.0 (R).