TB 9-6625-2270-35
Table 4. Ac Voltage - Continued
Test instrument
Calibrator output
Test instrument indications
range pushbutton
Voltage
Min
Max
2
1.9
V
50 kHz
1.8570
1.9430
20
19
V
20 Hz
18.570
19.430
20
19
V
1 kHz
18.570
19.430
19
V
10 kHz
18.570
19.430
20
20
19
V
50 kHz
18.570
19.430
200
190
V
20 Hz
185.70
194.30
200
190
V
1 kHz
185.70
194.30
200
190
V
10 kHz
185.70
194.30
200
190
V
50 kHz
185.70
194.30
750
V AC
750
V
40 Hz
730.0
770.0
750
V AC
750
V
1 kHz
730.0
770.0
750
V AC
750
V
10 kHz
730.0
770.0
750
V AC
200
V
50 kHz
191.0
209.0
10. dB Display
a. Performance Check
(1) Connect calibrator OUTPUT terminals to TI V and COMMON terminals.
(2) Simultaneously press both V and mA function pushbuttons. Then press 200mV
range pushbutton.
(3) Set TI range and calibrator for dBms and frequencies listed in table 5. TI will
indicate within the specified limits.
b. Adjustments. No adjustments can be made.
Calibrator
Test instrument indications
Test instrument
output
(dB)
range
Decibels
pushbutton
(dBm)
(kHz)
Min
Max
200 mV
-55
1
-54.50
-55.50
2
-20
1
-19.50
-20.50
2
-10
1
-9.50
-10.50
20
+15
1
+14.75
+15.25
200
+30
1
+29.75
+30.25
a. Performance Check
(1) Connect calibrator OUTPUT terminals to TI mA and COMMON terminals.