TB 9-6625-2271-50
Figure 6. Output level -
adjustment locations.
(8) Enter 2.048 V using DATA ENTRY keys and adjust A7R15 (fig. 6) for an ac
measurement standard indication of 1 mV greater than that recorded in (7) above (R).
(9) Enter 3.000 V using DATA ENTRY keys and adjust A6R23 (fig. 6) for an ac
measurement standard indication of 3.000 V (R).
a. Performance Check
(1) Connect TI OUTPUT HIGH to INPUT HIGH using a short shielded cable.
(2) Press keys and enter values using DATA ENTRY keys as listed in (a) through
(h) below:
(a) LCL INIT.
(b) SOURCE LEVEL.