TB 9-6625-2271-50
(c)
1 V.
(d)
SOURCE FREQ.
(e)
20 Hz.
(f)
ANALYZER DIST.
(g)
SPCL.
(h)
75 ENTER.
(3) If TI indication is not <.01 percent, perform b below.
(4) Press TI SOURCE FREQ key and enter 40 Hz using DATA ENTRY keys. If TI
indication is not <.01 percent, perform b below.
(5) Repeat (4) above using TI SOURCE FREQ settings listed in table 12.
Test instrument
SOURCE FREQ settings
(Hz)
80
160
320
640
1280
2560
5120
10240
19990
(7) Enter 21 kHz using DATA ENTRY keys. If TI indication is not <.02 percent,
perform b below.
(8) Enter 42 kHz using DATA ENTRY keys. If TI indication is not <.02 percent,
perform b below.
(9) Enter 84 kHz using DATA ENTRY keys. If TI indication is not <.056 percent,
perform b below.
(10) Enter 99 kHz using DATA ENTRY keys. If TI indication is not <.056 percent,
perform b below.
b. Adjustments. No
specific adjustments are provided for the output
circuitry of TI.
The adjustments listed in paragraph 12, Distortion Measurement Accuracy, could cause a
failure of this test. If distortion indications are only slightly high, perform paragraph 12b
then perform paragraphs 12a and 15.
16. SINAD
Measurement Accuracy
a. Performance Check