TB 9-6625-2273-40
NOTE
When you press ENTER pushbutton, front panel red RECAL
indicator will illuminate and remain illuminated until
adjustment process is completed. The number on the left of
display will increase one number each time ENTER
pushbutton is pressed, indicating calibration steps 1 through 12.
3 mV dc. Press ENTER
(9) Rotate TI control knob until multimeter indicates 0.0
pushbutton and observe TI readout displays 2 on left side.
l mV dc. Press ENTER
(10) Rotate TI control knob until multimeter indicates 0.0
pushbutton and observe TI readout displays 3 on left side.
3 mV dc. Press ENTER
(11) Rotate TI control knob until multimeter indicates 0.0
pushbutton and observe TI readout displays 4 on left side.
(12) Set multimeter to measure ac volts. Rotate TI control knob for a multimeter
indication between 1.9435 and 1.9455 V ac. Press ENTER pushbutton and observe TI
readout displays 5 on left side.
(13) Rotate TI control knob for a multimeter indication between 0.19435 and 0.19455 V ac.
Press ENTER pushbutton and observe TI readout displays a 6 on left side.
(14) Rotate TI control knob for a multimeter indication between 19.435 and 19.455 mV ac.
Press ENTER pushbutton and observe TI readout displays a 7 on left side.
(15) Rotate TI control knob for a multimeter indication between 1.9430 and 1.9460 V ac.
Press ENTER pushbutton and observe TI readout displays an 8 on left side.
(16) Rotate TI control knob for a multimeter indication between 0.19390 and 0.19500 V ac.
Press ENTER pushbutton and observe TI readout displays a 9 on left side.
(17) Rotate TI control knob for a multimeter indication between 0.19430 and 0.19460 V ac.
Press ENTER pushbutton and observe TI readout displays a 10 on left side.
(18) Rotate TI control knob for a multimeter indication between 19.390 and 19.500 mV ac.
Press ENTER pushbutton and observe TI readout displays an 11 on left side.
(19) Rotate TI control knob for a multimeter indication between 19.430 and 19.460 mV ac.
Press ENTER pushbutton and observe TI readout displays a 12 on left side.
(20) Rotate TI control knob for a digital multimeter indication between 1.9420 and
1.9470 mV ac. Press ENTER pushbutton.
(21) Press OUTPUT HEAD ON/OFF pushbutton (turn output OFF). Disconnect TI
from multimeter and set power module POWER switch to OFF. Remove TI from
power module.
(22) Set DIP switch SW3020 (rear panel of TI) toggle number 3 to open position
(pressed right). Insert TI into power module and set power module POWER switch to ON.
Allow 30 minutes for TI to warm-up then repeat a above.
14