TB 9-6625-2274-40
(3) Adjust UNITS/DIV dial to display 1 mA/D. Press AMPLITUDE MODE
NUMBER OF DIV (MULTIPLIER)
indicate between 4.9875 and 5.0125 mA.
(4) Adjust UNITS/DIV dial to display 5 mA/D. Press AMPLITUDE MODE
NUMBER OF DIV (MULTIPLIER)
indicate between 19.95 and 20.05 mA.
(5) Press OUTPUT ON pushbutton off.
(6) Remove multimeter and connect TI OUTPUT to digitizing oscilloscope
INPUT/TDR OUTPUT. Connect TRIGGER OUTPUT to digitizing oscilloscope TRIG.
(7) Position controls as listed in (a) through (g) below:
(a) Press AMPLITUDE MODE AMPL MODE CURRENT pushbutton.
(b) Press AMPLITUDE MODE FREQ
pushbutton to highlight 10 kHz.
p
(c) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
ushbutton to display X 4 DIV.
(d) Press TRIGGER ON push-button on.
(e) Press TRIGGER NORM pushbutton on.
(f) Adjust UNITS/DIV dial to display 1 mA/D.
(g) Press OUTPUT ON pushbutton on.
(8) Acquire signal using digitizing oscilloscope and adjust sensitivity to 1 mV/D.
Adjust digitizing oscilloscope offset to align top of square wave exactly on center horizontal
graticule.
(9) Disconnect digitizing oscilloscope TRIG input. Connect leveled sine wave
pushbutton to highlight +DC.
(10) Press AMPLITUDE MODE FREQ
(11) Set leveled sine wave generator for an output of 1 V, 10 kHz.
(12) Digitizing oscilloscope will display a signal indicating within .5 major division
of center horizontal graticule.
(13) Press OUTPUT ON pushbutton off and disconnect equipment setup.
b. Adjustments. See paragraph 6 d above.
10. Low Edge Amplitude Aberrations
a. Performance Check
(1) Connect TI OUTPUT to digitizing oscilloscope INPUT/TDR OUTPUT using
(2) Position controls as listed in (a) through (f) below:
(a) Press AMPLITUDE MODE AMPL MODE EDGE pushbutton.
p
(b) Press AMPLITUDE MODE NUMBER OF DIV (MULTIPLIER)
ushbutton to display X 6 DIV.
(c) Press AMPLITUDE MODE EDGE POLARITY pushbutton.
(d) Press AMPLITUDE MODE FREQ
pushbutton to highlight 100 kHz.
(e) Adjust UNITS/DIV dial to display 10 V/D.