SECTION I. IDENTIFICATION AND DESCRIPTION
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Programmable Calibration Generator Type Cg-5011 And Programmable Pulse Head Type 015-0611-00 Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . .
TB 9-6625-2274-40
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument
Performance specifications1
parameters
Programmable Calibration Generator, Tektronix, Type CG-5011
Range:  40 V p-p to 200 V p-p into 1M  2
Amplitude calibrator: Dc
A
40 V p-p to 5 V p-p into 50
and square wave
(0.25% of reading +1 V)
(chopped dc)
ccuracy:
Variable range:  9.9%
Frequency:  10 Hz, 100 Hz, 1 kHz, and 10 kHz
Current amplitude mode
Range:  1 to 100 mA
(0.25% of reading +2 A)
Accuracy:
Frequency:  10 Hz, 100 Hz, 1 kHz, and 10 kHz
Range:  Low amplitude output into 50  , 20 mV p-p to 1 V p-p
Edge amplitude mode
Risetime:  1.3 ns
Leading edge aberrations:  2% of square wave V p-p amplitude
Long term flatness4: Droop and tilt 0.5% maximum of pulse
amplitude after first 100 ns from leading
edge 50% amplitude point
Frequency:  10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz, and 1 MHz
Range:  High amplitude output into 1 M  , 1 V p-p to 100 V p-p3
Risetime:  100 ns
Leading edge aberrations: 2% of square wave V p-p amplitude
Long term flatness4: Droop and tilt 0.5% maximum of pulse
amplitude after first 500 ns from leading
edge 50% amplitude point
Frequency:  10 Hz, 100 Hz, 1 kHz, 10 kHz, and 100 kHz
9.9% from nominal
Variable amplitude:
Timing markers mode
Range:  0.5 ns to 5 s
Accuracy:
(0.005% of reading +10 ps)
Amplitude:  5 s to 2 ns l V p-p into 50
1 ns 350 mV p-p into 50
V
0.5 ns 100 mV p-p into 50
ariable marker: 9.9% of marker setting
T
Trigger output
Trigger amplitude: 1 V minimum into 50
rigger NORMAL rate:  5 s to 100 ns; remains at 100 ns for faster rates
Trigger mode RATE10:  Divides NORMAL rate by 10
Trigger mode RATE 100: Divides NORMAL rate by 100
Programmable Pulse Head, Tektronix, Type 015-0611-00
R
Fast edge amplitude
Range: 1.1 V 5% into 50
mode
isetime:  <150 ps
Leading edge aberrations: 3% pulse amplitude; not to exceed 40%, p-p
for adjacent peaks
Frequency:
100 Hz, 1 kHz, 10 kHz, and 100 kHz
Variable range: >l0%
1Specifications
are IAW MIS-35947 and may not agree with manufacturer's stated accuracies.
2200
V p-p check not verified due to standards limitation.
3100 V p-p check not verified due to standards limitation.
4Long term flatness not verified due to standards limitation.