TB 9-6625-2279-35
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Dc volts:
John Fluke, Model 5720A
Range: 242.5 mV to 1030 V
(p/o MIS-35947)
Accuracy: .75%
Dc current:
Range: 48.5 A to 515 mA
Accuracy: .75%
Ac volts:
Range: 2.375 to 1050 V
Accuracy: 1.25%
Range: 17.74Ω to 225531.9 Ω
Biddle-Gray, Model 71-631(7910328)
STANDARD
Accuracy: +2.83%
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the manufacturer's manual for this
TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance
of this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Performance Check
(1) Connect calibrator output to TI + and COM.
(2) Set TI function/range switch to 0.25 DCV.