TB 9-6625-2280-40
Table 2. Minimum Specifications of Equipment Required Continued.
Manufacturer and model
Common name
Minimum use specifications
(part number)
Agilent, Model 81150A Pulse
Amplitude: 5 V pk positive
Function Arbitrary Noise
Pulse width: 10 s
PRF: 10 kHz
Aeroflex, Model 2023B (2023B
OPT2/4/11/122)
Output power range: 0 to +3 dBm accuracy of
power meter reading
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in the procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraphs 8 through 12 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 12. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.